Accuracy Table

Transferpette

®

S, adjustable volume

!

20 °C

Ex

Final test values related to the nominal capacity (maximum volume)
indicated on the instrument, and the indicated volume steps obtained when
instrument and distilled water are equilibrated at ambient temperature
(20 °C/68 °F) and with smooth operation. According to DIN EN ISO 8655.

* A = Accuracy, CV = Coefficient of Variation

Volume range
µl

Volume step
µl

A*

± %

CV*

%

Increment
µl

Recommended
type of tip, µl

0.1 - 1

1

2

1.2

0.001

0.1 - 20

0.5

4

2.4

0.1

20

12

0.1 - 2.5

2.5

1.4

0.7

0.002

0.5 - 20

1.25

2.5

1.5

0.25

12

6

0.5 - 10

10

1

0.5

0.01

0.5 - 20

5

1.6

1

1

7

4

2 - 20

20

0.8

0.4

0.02

2 - 200

10

1.2

0.7

2

5

2

10 - 100

100

0.6

0.2

0.1

2 - 200

50

0.8

0.4

10

3

1

20 - 200

200

0.6

0.2

0.2

2 - 200

100

0.8

0.3

20

3

0.6

100 - 1000

1000

0.6

0.2

1

50 - 1000

500

0.8

0.3

100

3

0.6

500 - 5000

5000

0.6

0.2

5

500 - 5000

2500

0.8

0.3

500

3

0.6

1000 - 10000

10000

0.6

0.2

10

1000 - 10000

5000

0.8

0.3

1000

3

0.6