Resetting the Controller for Use With Heating Mantles.

If you want to use your controller with heating mantles after it’s been tuned for a different style heater,
rather than autotuning the controller with the heating mantle, J-KEM recommends that the controller be
manually tuned by following the procedure below.

Procedure 1

.

Perform when using heating mantles with round bottom flasks.

[This procedure takes about 2 minutes to perform]

1.

Press and hold in both the ▼ and ▲ keys on the front of the temperature meter until the word “tunE” appears in the display, then release
both keys.

2.

Press the ▲ key once and the word “bAnd” will appear in the display.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the ▼ or ▲ key until the value “10” appears in the display, then let go of
all the keys.

3.

Press the ▲ key once and the word “int.t” will appear in the display.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the ▼ or ▲ key until the value “10” appears in the display, then let go of
all the keys.

4.

Press the ▲ key once and the word “dEr.t” will appear in the display.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the ▼ or ▲ key until the value “50” appears in the display, then let go of
all the keys.

5.

Press the ▲ key once and the word “dAC” will appear in the display.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the ▼ or ▲ key until the value “3.0” appears in the display, then let go of
all the keys.

6.

Press the ▲ key once and the word “CyC.t” will appear in the display.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the ▼ or ▲ key until the value “30” appears in the display, then let go of
all the keys.

7.

Press and hold in both the ▼ and ▲ keys until the temperature appears in the display, then release both keys.


3.11

Changing the Temperature Display Resolution The controller is programmed to display temperature

with 0.1

o

C resolution. The controller can also display 1

o

C resolution by following the procedure below (the

display can also be changed to read in

o

F, call for information). There are two reasons to change the display

resolution:

1) To enter a setpoint faster (the display scrolls 10X faster in 1

o

mode than in 0.1

o

mode).

2) To display temperatures above 999.9

o

.

1.

Press and hold in both the Ð and Ï keys on the front of the temperature meter until the word “tunE” appears in the display, then release
both keys.

2. Press

the

Ð key once and “LEVL” appears in the display.

First hold in the ‘*’ key, then while holding in the ‘*’ key press the Ï key until “2” appears in the display then let go of all

the

keys.

3. Press

the

Ï key repeatedly until the word “diSP” appears in the display.

First hold in the ‘*’ key, then while holding in the ‘*’ key press the Ð or Ï key until the value “1” [not “0.1”] appears in

the display, then let go of all the keys.

4.

Press and hold in both the Ð and Ï keys until the temperature appears in the display, then release both keys.



3.12 Fusing. The controller is protected by a bank of fuses designed for solid-state equipment that are inside

the controller. None of the fuses are user serviceable and should be replaced by a qualified electrician If
fuses are replaced THEY MUST BE REPLACED BY AN EXACT EQUIVALENTS (such as a
Littlefuse #L15S 15). Fuses can be obtained from electrical supply houses or from J-KEM.

Advertising