Maintenance, Leaning, Technical data – Chemglass CG-2024-10 User Manual

Page 3: Wiring diagram

Advertising
background image

3

Over Curr

Rpm error

This error message is displayed if the chuck is accidentally blocked or if the number of revs exceeds the permitted value.
Over Temp

Temperature error

This error message may appear if the unit is used for heavy-duty work for prolonged periods of time and at an ambient
temperature that exceeds the max. permitted temperature. Before resetting the instrument allow it to cool down.
Over Volt

Voltage surge

The software limits the power supply of the motor to the nominal value of the motor itself (60V). A high speed setting or a
heavy load on the chuck may prompt a higher voltage. In this case the voltage is limited to the nominal value of the motor
(60V) and the number of revs is reduced accordingly. In these conditions the following symbol appears in the top rh corner

of the display

. If for any reason the power supply of the motor exceeds the maximum threshold value of 70V,

“OVERLOAD” will appear on the display and the unit will shut down.
NOTE: To reset the unit remove the load from the chuck and turn the unit off and on again using the mains switch.

4. Maintenance

No routine or extraordinary maintenance is necessary apart from periodically cleaning the unit as described in this manual.
In compliance with the product guarantee law, repairs to our units must be carried out in our factory, unless previously
agreed otherwise with local distributors.

4.1 Cleaning
Disconnect the unit from the power supply and use a cloth dampened with a mild, non-flammable detergent.

5. Technical data

6. Wiring diagram

Admitted power supply

From 80 to 260Volt 50/60Hz

Max. input / output power

130W / 120 W

Dimensions (WxHxD)

80x230x196 mm (3.1x9.0x7.7 in)

Weight

3 Kg (6,6 lbs)

Speed range at nominal load

50

– 2000rpm

Speed indicator / resolution / accuracy

LCD / 1rpm / ±2 rpm

Max torque stirrer shaft / accuracy

80Ncm / ±5 Ncm

Timer with automatic cut-off

From 000’:00” to 999’:59”

Max stirring quantity H

2

O

40 litres

Clamping chuck clamping range

From 1 to 10mm

Overload and overtemperature protection

Automatic cut-off

Enviromental / Storage ambient temperature

5 - 40°C (41 - 104°F) / -

10… + 60°C (14 - 140°F)

Max humidity

80%

Electr. safety level CEI-EN60529

IP40

Pollution degree / installation cat. CEI-EN61010-1

2

Max altitude installation

4000m (13123ft) asl

1. Fuse 5x20mm, 2x3A
2. Electric motor
3. Electronic board
4. Encoder
5. ON-OFF switch
6. Network filter

Advertising