Do not use damaged components. Safe operation is only

guaranteed with the accessories listed in the “Accessories”

section. Accessories must be securely attached to the

device and must be removed when not in use. Always

disconnect the power before fitting accessories.

• When using the external temperature sensor, the tip of the

measuring sensor must be at least 5-10mm from vessel’s

bottom and wall.

• The instrument can only be fully disconnected from the

main power supply by turning off the main or disconnecting

the plug.

• The voltage stated on the label must correspond to the

main power supply.

• Ensure that the main power supply cable does not touch

the hotplate. Do not cover the device.

• The instrument must only be operated by adults.

• Keep away from high magnetic fields.

2 Proper Use

The instrument is designed for mixing and/or heating

liquids in laboratories:

Do not use more than 2000 meters above sea level

Temperature between 0℃and 40℃

Installation Type: Product is designed to connect the

indoor outlet, where voltage fluctuation is not more

than ±10% compared with normal.

Observe the minimum distances between multiple

units, and distances to the rear wall and above the

assembly (min. 100 mm).

Figure 1

This device is not suitable for using in residential areas or

other constraints mentioned in Chapter 1.

>100mm

>100mm