Specifications – Chemglass AF-0304 User Manual

Page 7

Advertising
background image

2. SPECIFICATIONS

5

• Pressure Range: 0.01 to 100.00 mTorr, 1 to 2000

mTorr, 0.01 to 20.00 Torr, depending on designed

range of instrument.

• Calibration: Air or Nitrogen.

• Accuracy: Better than 5% of value or:

(for 100 mTorr gauges)

±0.03 mTorr,

(for 2000 mTorr gauges)

±3 mTorr,

(for 20 Torr gauges)

±0.03 Torr,
whichever is larger.

• Set Point Range: Full scale, equal to full scale range

of instrument.

Set Point Repeatability: 2 digits.

• Analog Output: 0 to 2.000 volts = 0-2000 mTorr into

2000 ohm load or higher; or 2.000 volts = 20.00 Torr;
or 1.0000 volt = 100 mTorr. 2 pin connector included.

• Gauge Tube: Gauge tubes for 2000 mT and 20T

have 1/8 inch NPT male thread or fit into quick
connects sized for 1/8 inch pipe (0.405 inch dia.)
100 mTorr tube has 3/4 inch O.D. smooth tubulation.

• Power Consumption: Approx. 1 Watt for indicator,

2 Watts for Controller @ 115 VAC (4 Watts for 20 Torr
models).

• Relays: SPDT 3 Amp @115VAC; non-inductive loads.

• Temperature Effects: (For 2000 mT): changes in

ambient temperature between 0°C and 50°C
change the reading less than 0.2 mT per °C at 50
mT; less than 4 mT per °C at 900 mT. (For 20 Torr):
Changes in ambient temperature between 0°C and
50 °C change the reading less than 0.01 Torr per °C.
(For 0.01 to 100 mT): Changes in ambient temperature
between 0° C and 50° C change the reading less than
0.004 mTorr per °C at hard vacuum and 0.4 mTorr at
50 mTorr.

• Line Power: 115VAC, 50/60 Hz; 230 VAC optional at

no additional cost.