WIKA FLR User Manual

Page 7

Advertising
background image

GB

WIKA operating instructions level sensors model RMG

7

13442163.01 06/2010 GB/D/F

6.

Function test

6. Function test

A function test can only be carried out when the level sensor has been

removed.

■

Disconnect the cable

■

Connect an ohmmeter to two wires

■

Move the float by hand from Min. to Max.

■

The displayed resistance value changes continuously depending on

the connected wire colours (Tab. 1):

black-brown (R1)

blue-brown (R2)

black-blue (Ri)

Resistance value incre-
ases in proportion to
the height of the float.

Resistance value drops
from the value of the
total resistance inverse-
ly proportionally to the
height of the float.

Indication of the total
resistance (Ri)

Table 1

The value of the resistance chain is given on the tag plate mounted at

the level transmitter.

Advertising