Dispensing antibodies, Multi-dispense, Run or save program – Next Advance DX User Manual

Page 7: Perator, Anual for the, Reedom, Ocker

Advertising
background image

O

PERATOR

S

M

ANUAL FOR THE

F

REEDOM

R

OCKER

DX

Table 1. Sample Program - overnight Western antibody staining and washing

7

DRAIN

Description

FROM

VOLUME

TIME

SPEED

TO

(1 to 9)

(yes or no)

(white, blue,

y'low)

(0 to 600 ml)

(hr:min, from
:00 to 90:00)

(30 to 100)

(red, black, or

green)

1

100

2:00

30

red

blocker in trays

2

3

yes

white

100

:20

80

red

3X 20 min wash

3

1

yes

blue

30

7:00

30

black

10 mL Ab (use 50 mL tube,

clearing rinse) and recover

4

3

no

white

100

:05

100

red

3X 5 min wash

5

2

yes

white

200

:30

80

red

2x 30 min wash

6

1

no

yellow

50

1:30

30

green

50 mL secondary Ab,

recover

7

3

yes

white

100

:30

80

red

3X 30 min wash

8

not used

9

not used

ROCK

No. of Steps (up to 9)

STEP

REPS

CLEARING

RINSE

FILL

rinses. For example, to do a series
of 3 rinses, set the number of
repetitions to 3.

Dispensing Antibodies
To dispense small quantities, e.g. 10
mL of antibodies or other precious
reagents, fill up the 50 mL tube as
the source container, loosen the
cap, and set the fill volume to 30 mL
and set the clearing rinse on. The
clearing rinse is preset to empty the
tubing of the last reagent/buffer.
Setting the dispense to 30 mL
makes sure that enough air is
pumped through the tubing after the
50 mL of antibody, so that the
antibody is pumped through the
tubing and into the trays.
For example, to dispense 10 mL to
each of the trays, fill the source
container with 50 mL and set the fill
volume to 30 ml. The clearing rinse
will use up approximately 4 mL (not
recoverable). The rest of the
antibody solution will then be

dispensed into the trays. Allow a
little extra since some trays might
receive 1 mL more than others.

Multi-Dispense
In multi-dispense models, the four
tubes which sit atop the dispense
heads are the tubes which you can
fill to allow multi-dispense. To fill
from these tubes, simply select
“tubes” as the fill source. The entire
volume contained in these tubes will
be dispensed; up to 20mL (do not
overfill the tubes as the additional
volume may not be dispensed).

Please note that the clearing rinse
option will be automatically set to
off and the volume selection will be
disabled when the tubes are
selected as the fill source.

Run or Save Program
Once you have set the parameters
for all the steps you want in your
program you can save the program
or just run it.

7

Advertising