Next Advance SP510 User Manual

Next Advance Pumps

Advertising
background image

Next Advance

Model: SP1010, SP510 syringe pump

Maximum and minimum flow rates

Manufacturer

Syringe
(cc)

Inside
Diameter

(mm)

Maximum
Rate

(mL/hr)

Minimum
Rate

(µl/hr)

Maximum
Rate

(mL/mn)

B-D

1

4.699

191.1

1.459

3.185

3

8.585

637.9

4.868

10.63

5

11.99

1244

9.495

20.74

10

14.43

1802

13.76

30.04

20

19.05

3141

23.97

52.35

30

21.59

4035

30.79

67.25

60

26.59

6120

46.7

102

Monoject

1

5.74

285.2

2.176

4.753

3

8.941

692

5.28

11.53

6

12.7

1396

10.66

23.26

12

15.72

2139

16.33

35.65

20

20.12

3504

26.74

58.4

35

23.52

4788

36.54

79.81

60

26.64

6143

46.88

102.3

Terumo

1

4.7

191.2

1.459

3.187

3

8.95

693.4

5.291

11.55

5

13

1462

11.17

24.38

10

15.8

2160

16.49

36.01

20

20.15

3514

26.82

58.57

30

23.1

4619

35.25

76.98

60

29.7

7635

58.26

127.2

Air-Tite

10

15.9

2188

16.7

36.47

20

20.25

3549

27.09

59.16

30

22.5

4382

33.44

73.03

50

29

7280

55.55

121.3

Syringe
(uL)

Inside
Diameter

(mm)

Maximum
Rate

(ul/hr)

Minimum
Rate

(µl/hr)

Hamilton
Microliter

0.5

0.103

91.83

0.001

1

0.146

184.5

0.002

2

0.206

367.3

0.003

5

0.326

919.9

0.008

SGE

0.5

0.1

86.56

0.001

1

0.15

194.7

0.002

Hamilton

Micro Liter

10 ul

0.48

1.994

0.016

0.9

SGE

500 ul

3.26

Syringe

(mL)

Inside

Diameter

(mm)

Maximum

Rate

(mL/mn)

Minimum

Rate

(µl/hr)

Air-Tite

140

38

208.3

95.37

Advertising
This manual is related to the following products: