RADWAG WTB 2000 Precision Balance User Manual

Page 52

Advertising
background image

• Place the required weight on the pan,

• Calibration process starts automatically after placing the adequate

weight that is signalled by the following message:

• The completion of the calibration procedure is signalled by the

following message:

• Take off the weight from the pan, the message <donE> is displayed

for 1s and the program returns to the calibration submenu:

• Calibration process can be terminated anytime by pressing

which is signalled by the following message on the display:

• Return to weighing with saving changes that have been made.


Caution:
If the calibration process (span adjustment) lasts longer than 15 the

<Err8>
message will be displayed and short audible signal will be

emitted. Press

to perform calibration again with more stable

ambient conditions!

- 52 -