Formatting rechargeable battery packs – RADWAG Scales of WPT series User Manual

Page 42

Advertising
background image

• Message

<nImh> appears on the display (for indicators in

plastic casings) or <SLA> (for indicators in metal housings).

• In case of damaging accumulators or lack of it the pictogram

blinks quickly (period about 0.5 sec).

Notice:
Indicators in plastic casings are equipped with the set of rechargeable
batteries NiMH R6 (AA) and power adapter.

Procedure:

• Enter the submenu <P5.othr> and then:

CHr6 YES

- enabled

CHr6 no

- disabled

Return to weighing:

See 10.2.2.



14.4.4. Formatting rechargeable battery packs

Every plastic indicator is equipped with a brand new NiMH R6 (AA)
battery pack and a power adapter. They need formatting after first
powering up. It is crucial for batteries lifetime to undertake this
process. Formatting consist in charging and total discharging
(without meantime charging).

- 42 -