Send the stable result in basic unit, Send the result immediately in basic unit, In basic unit – RADWAG Scales of WPT series User Manual

Page 68: Send the result immediately, To_tara, Command executed frame format, Syntax: s cr lf, Possible answers, Frame format, Syntax: si cr lf

Advertising
background image

- 68 -

TO_TARA

CR LF

- command executed


Frame format:

1 2

3

4

5-6 7-15

16 17 18 19 20 21

T O space stability space tare space

unit

CR LF

Tare - 9 characters with decimal point justified to the right
Unit

- 3 characters justified to the left



19.4.4. Send the stable result in basic unit

Syntax: S CR LF

Possible answers:

S_A CR LF
S_E CR LF

- command accepted and in progress
- time limit for stable result exceeded

S_I CR LF

- command comprehended but cannot be executed

S_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in basic unit is returned


Frame format:

1 2-3

4

5 6 7-15 16 17

18

19

20

21

S space stability space sign

mass

space unit CR

LF

Example:

S CR LF – computer command
S _ A CR LF - command accepted and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.


19.4.5. Send the result immediately in basic unit

Syntax: SI CR LF

Possible answers: