Error commands – RADWAG Scales of WPT series User Manual

Page 74

Advertising
background image

- 74 -

20. ERROR COMMANDS

Err2

- Value beyond the zero range

Err3

- Value beyond the tare range

Err4

- Calibration mass or start mass beyond the acceptable

range (

±1% for weight, ±10 for start mass)

Err5

- Mass of a single piece lower than the scale division

Err8

- Exceeded the time for tarring, zeroing, start mass

adjustment or span adjustment

NULL

- Zero value from the AD converter

FULL2

- Measurement

range overflow

LH

- Start mass error, the mass on the weighing platform is

beyond the acceptable range (-5% to +15% of start mass)

5-FULL

- Display range overflow in totalizing


Notice:

1. Errors: Err2, Err3, Err4, Err5, Err8, null, that appear on the display

are also signalled by a short beep sound (about 1 sec.);

2. Error FULL2 that appears on the display is also signalled by

a continuous sound until the cause of error disappears.