Sense:function, Sense:sweep:offset:points, Sense:sweep:points – Agilent Technologies 6631B User Manual

Page 53: Sense:sweep:tinterval

Advertising
background image

Language Dictionary - 4

53

SENSe:FUNCtion

Agilent 66312A, 66332A Only

This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.

Command Syntax

SENSe:FUNCtion <function

Parameters

"VOLTage" | "CURRent"

Examples

SENS:FUNC "VOLT"

Query Syntax

SENSe:FUNCtion?

Returned Parameters

<SRD>

SENSe:SWEep:OFFSet:POINts

Agilent 66312A, 66332A Only

This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger.

Command Syntax

SENSe:SWEep:OFFSet:POINts <NRf+>

Parameters

-4095 through 2,000,000,000

*RST Value

0

Examples

SENS:SWE:OFFS:POIN -2047

Query Syntax

SENSe:SWEep:OFFSet:POINts?

Returned Parameters

<NR3>

Related Commands

SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR

SENSe:SWEep:POINts

This command defines the number of points in a data sweep.

Command Syntax

SENSe:SWEep:POINts<NRf+>

Parameters

0 through 4096

*RST Value

2048

Examples

SENS:SWE:POIN 1024

Query Syntax

SENSe:SWEep:POINts?

Returned Parameters

<NR3>

Related Commands

SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR

SENSe:SWEep:TINTerval

This command defines the time period between samples

Command Syntax

SENSe:SWEep:TINTerval<NRf+>

Parameters

15.6 microseconds through 31200 seconds

*RST Value

15.6 microseconds

Examples

SENS:SWE:TINT 31.2E-6

Query Syntax

SENSe:SWEep:TINTerval?

Returned Parameters

<NR3>

Related Commands

SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR

Advertising