Troubleshooting (1) 11. troubleshooting (2), English – AEG BMG 5611 User Manual

Page 92

Advertising
background image

English

92

10. TROUBLESHOOTING (1)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display shows
abnormal result

The cuff position was not
correct or it was not properly
tightened

Apply the cuff correctly and try
again

Body posture was not correct
during testing

Review the section “BODY POS-
TURE DURING MEASUREMENT”
of the instructions and try again.

Speaking, arm or body
movement, angry, excited or
nervous during testing

Re-test when calm and without
speaking or moving during the test

Irregular heartbeat (arrhyth-
mia)

It is inappropriate for people with
serious arrhythmia to use this
Electronic Sphygmomanometer.

11. TROUBLESHOOTING (2)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD shows the sym-
bol for a low battery Low battery

Change the batteries

LCD shows “Er 0”

Pressure system is unstable
before measurement

Don’t move and try again.

LCD shows “Er 1”

Fail to detect systolic pressure

LCD shows “Er 2”

Fail to detect diastolic pres-
sure

LCD shows “Er 3”

Pneumatic system blocked or
cuff is too tight during inflation

Apply the cuff correctly and try
again

LCD shows “Er 4”

Pneumatic system leakage or
cuff is too loose during infla-
tion