Analog I/O Modules Calibration


6-6 ADAM 5000 IO modules User’s Manual

Calibration voltage (ADAM-5017H)

Module

Input Range Code
(Hex)

Input Range

Span Calibration Voltage

00h

±10 V

+10 V

01h

0 ~ 10 V

+10 V

02h

±5 V

+5 V

03h

0 ~ 5 V

+5 V

04h

±2.5 V

+2.5 V

05h

0 ~ 2.5 V

+2.5 V

06h

±1 V

+1 V

07h

0 ~ 1 V

+1 V

08h

±500 mV

+500 mV

09h

0 ~ 500 mV

+500 mV

0ah

4 ~ 20 mA

*(1)

5017H

0bh

0 ~ 20 mA

*(1)

Table 6-2: Calibration voltage of ADAM-5017H

Note: You can substitute 2.5 V for 20 mA if you remove the current conversion

resistor for that channel. However, the calibration accuracy will be limited
to 0.1% due to the resistor's tolerance.