Medical administrator tab – Barco NIO E-2620 User Manual

Page 97

Advertising
background image

Application settings

Nio® medical display systems

97

Figure 50: Dual-head equalization

The reference display is the first calibrated display in the system.

The Equalization option will be available only if:

•

the system contains at least one display of the same type as
the one being calibrated

•

at least one of the displays of the same type has already
been calibrated before using a sensor.

MediCal Administrator tab

Figure 51: MediCal Administrator tab

Here you can connect to MediCal Administrator, if present.

To connect to MediCal Administrator, click the MediCal
Administrator
button. As a result, the MediCal Administrator

BARCO

BARCO

c

"

(1) Reference display
(2) Calibrated display: Luminance automatically matched
to reference display