Bacharach 24-7224 User Manual

Introduction

Advertising
background image

Introduction

The Bacharach Flue-Gas Sample Conditioning System consists of the

following components:
• Probe with Peltier Cooler

• Condensate Pump with Water Trap

• Power Supply
The function of this system is to remove water vapor from the flue-gas

sample, thus preventing the formation of water droplets inside the

probe hose. If water droplets were to form inside the hose, a portion of

the gas sample could be absorbed by the water, thus resulting in lower-

than-actual readings of NO

2

and SO

2

.

Water vapor is removed from the flue-gas sample by passing the sample

through a Peltier Cooler where the sample is chilled, which in turn,

causes the water vapor contained in the gas sample to be removed from

the sample.
The water that was extracted from the gas sample is then pumped out

of the Peltier Cooler and into a water trap.
The now dry, conditioned flue-gas sample is then sent to its associated

combustion test equipment for analysis.

Product Leadership • Training • Service • Reliability

Figure 1. Flue-Gas Sample Conditioning System

Probe w/ Peltier Cooler

Power Supply

Instruction 24-9415

Flue-Gas Sample

Conditioning System

P/N 24-7224

Rev. 5 – June 2005

Pump
w/ Water Trap