Philips TensRelief 12 selbstklebende Elektroden User Manual
Page 42

Problem
Possible cause Solution
I
experienced
an
unpleasant
sensation
when I
touched the
connectors
of the TENS
units or the
electrodes.
You did not
decrease the
intensity to
0 to end the
treatment
before you
removed the
TENS units.
Always decrease the intensity to 0
mA before you remove the TENS
units. In this way you avoid the
unpleasant sensation when you touch
the connectors of the TENS units
or the extension cable, or touch the
electrodes. This sensation is unpleasant
but not harmful.
The
stimulation
feels
different or
less pleasant
than usual.
The intensity
is too high or
too low.
Press the + or - side of the intensity
button to increase or decrease the
intensity.
The electrodes
are not in the
right place.
Switch off the TENS units and detach
them from the electrodes. Remove
the electrodes from the skin and move
them to a slightly different place.
The
electrodes are
not properly
attached.
Switch off the TENS units and detach
them from the electrodes.Check if the
electrodes are attached to the skin
properly.
The electrodes
are too dirty.
You can try to clean the electrodes
(see chapter ‘Cleaning’). If this does not
help, replace the electrodes.
The electrodes
are worn.
Replace the electrodes.