Philips TensRelief 12 selbstklebende Elektroden User Manual

Page 42

Advertising
background image

Problem

Possible cause Solution

I

experienced

an

unpleasant

sensation

when I

touched the

connectors

of the TENS

units or the

electrodes.

You did not

decrease the

intensity to

0 to end the

treatment

before you

removed the

TENS units.

Always decrease the intensity to 0

mA before you remove the TENS

units. In this way you avoid the

unpleasant sensation when you touch

the connectors of the TENS units

or the extension cable, or touch the

electrodes. This sensation is unpleasant

but not harmful.

The

stimulation

feels

different or

less pleasant

than usual.

The intensity

is too high or

too low.

Press the + or - side of the intensity

button to increase or decrease the

intensity.

The electrodes

are not in the

right place.

Switch off the TENS units and detach

them from the electrodes. Remove

the electrodes from the skin and move

them to a slightly different place.

The

electrodes are

not properly

attached.

Switch off the TENS units and detach

them from the electrodes.Check if the

electrodes are attached to the skin

properly.

The electrodes

are too dirty.

You can try to clean the electrodes

(see chapter ‘Cleaning’). If this does not

help, replace the electrodes.

The electrodes

are worn.

Replace the electrodes.

English

42