5.3.6 If Fuse blows, connect the unit through a variable AC Source with a current meter

and monitor the current at no load. If the current is very high (more than 2 ampere
or so ) load, check for Bridge Rectifier short or shorted Diode, or Secondary
Winding short, etc

5.3.7 If output voltage or current DPM's are not reading, check for loose connections

especially in crimping, soldering, of Connectors.