CL17 Free and Total Chlorine Calibration/Verification

Page 7

5449089Cal Kit.fm

CL17 Free and Total Chlorine Calibration/Verification

Note: Take care when
preparing the standard.
An incorrectly prepared
standard can skew the
calibration curve and may
result in a calibration failure.

6.

Using the concentration of the ampule printed on the box, calculate the expected
standard concentration using the following equation:

7.

Write the calculated concentration on the bottle.

Running the Chlorine Standard

1.

Perform steps

1

15

of

Running the Zero Standard

using the prepared chlorine

standard in place of the Zero Standard (Organic-free Water) and in step

12

, use the

exception below:

New Style Analyzer:

a.

After the four measurements are complete, press the

MENU

key.

b. Record each reading if the analyzer is not connected to a computer.

c.

Using the up and down arrow keys, scroll to SETUP and press

ENTER

.

d. Scroll to CAL STD and press

ENTER

.

e.

Using the up and down arrow keys, enter the calculated concentration and
press

ENTER

.

f.

Press

ENTER

again to set the slope to its new value.

g. Exit to main display.

Old Style Analyzer:

a.

After the four measurements are complete, use the numeric keypad to enter the
calculated concentration.

b. Press the

STD

key.

c.

Exit to main display.

Calculating Average Concentration, %RSD and %Difference

1.

Calculate the average concentration of the standard (omit the first replicate) by either
manually calculating it or by using a spreadsheet program.

where a, b and c are the actual concentrations for the 2nd, 3rd and 4th readings of the standard.

2.

Calculate the standard deviation (SD) and the % relative standard deviation (%RSD).
The %RSD value should be less than 3%.

Important Note: If %RSD value is greater than 3%, rerun the test with fresh standard reagents.

Concentration of Ampule

(

)

0.020 L

(

)

×

0.52 L

------------------------------------------------------------------------------------------------------

Standard Concentration

=

Avg

a

b

c

+

+

3

----------------------

=

S D

a

avg

(

)

2

b

avg

(

)

2

c

avg

(

)

2

+

+

(

)

1/2

3

------------------------------------------------------------------------------------------------------------

=

%RSD

SD

100

×