Analog interface

Pin 1, +24 V

DC

, 0.5 A

Auxiliary voltage for peripherals
(max. current 0.5A)

Pin 2, status error
This pin is connected to 'status 0'
(pin 9) if the pump is switched off
or in the case of an error.
max.: AC 70 V / 1 A
DC 48 V / 1 A

Pin 3, output 0
Reference point for the 'busy'
output, potential-free.

Pin 4, busy
See page 23

Pin 5, direction
If this input is set to high (5, 12 or
24 V) relative to 'GND' (Pin 8) the
pump rotates in the counter-
clockwise direction.
Configuration see page 17

Pin 7, speed IN
For external speed control
0 – 5 V (1M

Ω)

, 0 – 10*V (20k

Ω)

0 – 20 mA (250

Ω)

, 4 – 20 mA (250

Ω)

Configuration see page 17

Pin 8, GND (ground)
Reference potential for all analog
and digital inputs and the analog
output. GND lies on earth
potential.

Pin 9, status 0
Reference potential for both
status outputs, potential-free.

Analogschnittstelle

Pin 1, +24 V

DC

, 0.5 A

Hilfsspannung für Peripheriegerä-
te (Belastbarkeit: 0.5 A).

Pin 2, status error
Bei abgeschalteter Pumpe oder
bei einem Fehler ist dieser Aus-
gang intern mit dem Anschluss
'status 0' (Pin 9) verbunden.
max.: AC 70 V / 1 A
DC 48 V / 1 A

Pin 3, output 0
Bezugspunkt für den 'busy'-Aus-
gang, potentialfrei.

Pin 4, busy
Siehe Seite 23

Pin 5, direction
Wird dieser Eingang gegenüber
'GND' (Pin 8) auf High (5, 12 oder
24 V) gelegt, dreht die Pumpe im
Gegen-Uhrzeigersinn. Konfigurati-
on siehe Seite 17

Pin 7, speed IN
Für externe Drehzahlsteuerung
0 – 5 V (1M

Ω)

, 0 – 10*V (20k

Ω)

0 – 20 mA (250

Ω)

, 4 – 20 mA (250

Ω)

Konfiguration siehe Seite 17

Pin 8, GND (Masse)
Bezugspunkt für alle analogen
und digitalen Eingänge, sowie
den analogen Ausgang. GND liegt
auf Erdpotential.

Pin 9, status 0
Bezugspunkt für die beiden Sta-
tus-Ausgänge, potentialfrei.

Digitale Eingänge (5,12 oder 24

*

V

DC

)

Digital inputs (5,12 or 24* V

DC

)

Entrées numériques (5,12 ou 24* V

CC

)

Pin 5, direction

Pin 13, start

Digitale Ausgänge (Relais)

Digital outputs (Relais)

Sorties numériques (Relais)

Pin 2 / 9, error

Pin 10 / 9, ok

Pin 4/11/12, busy

Analog-Eingang

Analog input

Entrée analogique

Pin 7, speed IN

0 – 5 V

DC

/ 0 – 10 V

DC

* / 0 – 20 mA / 4 – 20 mA

Analog-Ausgang

Analog output

Sortie analogique

Pin 17, speed OUT, 0 – 5 V

DC

/ 0 – 10 V

DC

Pin 1, +24 V

DC

0.5 A

* = Default Wert
* = default value

* = valeur par défaut

Interface analogique

Pin 1, +24 V

CC

, 0.5 A

Tension auxiliaire pour appareils
périphériques (courant maximal
0.5 A)

Pin 2, status error (erreur d'état)
Lorsque la pompe est éteinte ou
en cas d'erreur, cette sortie est
reliée au raccordement «status 0»
(pin 9) max. : CA 70 V / 1 A

CC 48 V / 1 A

Pin 3, output 0 (sortie 0)
Point de référence pour la sortie
'busy' (active), sans potentiel

Pin 4, busy
Voir page 23

Pin 5, direction (sens)
L'entrée du Pin 5 doit avoir une
différence de potentiel de 5, 12
ou 24 V par rapport à 'GND'
(PIN 8). La pompe tourne dans le
sens inverse des aiguillles d'une
montre.
Configuration voir page 17

Pin 7, speed IN (entrée vitesse)
Pour la commande externe de la
vitesse
0 – 5 V (1M

Ω)

, 0 – 10*V (20k

Ω)

0 – 20 mA (250

Ω)

, 4 – 20 mA (250

Ω)

Configuration voir page 17

Pin 8, GND (ground) (masse)
Point de référence pour toutes les
entrées analogiques et
numériques et pour la sortie ana-
logique. GND est sur le potentiel
de terre.

Pin 9, status 0 (état 0)
Point de référence pour les deux
sorties d'état, sans potentiel.