BNC 1120 - 1pS Time Interval Counter User Manual

Page 16

Advertising
background image

Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-16 / 36 -

3.2.4. Default value

The default values are the following:

At factory:

@IP: 99.0.0.18,

@GW: 99.0.0.1,

Mask: 255.0.0.0,

At start up:

START: OFF,

STOP: OFF,

Level: 0.5 V,

3.3.

Rear panel operation

3.3.1.1. Rear panel connectors

10MHZ input (BNC type): When the Rack is in external clock mode the internal oscillator
will phase lock to a 10MHz signal applied to this connector

10MHZ output (BNC type): This signal is a 10 MHz square wave. It is a copy of the
internal oscillator and can be used to verify the frequency or to drive another BNC 1120,
which are operating, in external mode.

Option: Optical input (SC/PC type): This optical input is compliant to GFT3001 for
timing system measurement. In that mode the BNC 1120 use the clock and the
triggers provided by the optical network connected to GFT3001.

Power source: For 110 to 220 Voltages AC line.

3.3.1.2. Rear panel control

Power switch: Turns AC power on.