Faults and error codes – IKA T 65 digital ULTRA-TURRAX User Manual

Page 22

Advertising
background image

22

Faults and error codes

Faults on the T 65 basic ULTRA-TURRAX

®

Cause

Effect

Solution


Power supply failure

The unit cannot be switched on at

Checking the power supply should be perfor-

the main switch.

med by authorised personnel.

The equipment has switched itself off.

Precautionary measure:

Check the power supply

Stand not in position (if the

The unit cannot be switched on at

Move the stand into position.

stand safety limit switch is

the main switch.

connected)

The equipment has switched itself off.

Precautionary measure:

Check the position of the stand

Overload

The overcurrent limit switch has

Remedy the overload, e.g.:

switched the unit off.

- the viscosity of medium is too high

- Solids jamming the stator/rotor

- Dispersing element defective (blocked).

If none of these causes was the cause of the problem, a fault other than those described must have occurred and the unit must be
checked by an authorised person.

Error codes T 65 digital ULTRA-TURRAX

®

If a fault occurs with the unit, proceed as follows:

Switch off the unit at the main switch (A)

Carry out corrective measures

Restart the device

All error codes are displayed with the legend "trip", i.e. the word "trip" is displayed alternating with the relevant error code.

Error

Cause

Effect

Solution

code
lt.AC Motor overload

Device off

Reduce the load by reducing the speed.

Overloads are especially liable to occur with

high-viscosity media and at high speeds.

Oht1 Overtemperature

Device off

Switch the unit off and allow it to cool down,

try reducing the load or the operating cycle.

OI.AC Overcurrent at the frequency

Device off

Reduce the load by reducing the speed. If this

inverter

error appears even when idling with no load,

the frequency inverter is defective. Consult

the Service Engineer.

OU

Overvoltage at the frequency

Device off

Check the mains power cable values and the

inverter

power supply.

Note: High-frequency noises during operation of the T 65 digital ULTRA-TURRAX

®

may to some extent originate from the frequency

inverter, depending on the design. They certainly do not indicate a defect within the unit.

Note: The green and red LEDs around the "Speed setting" rotary knob (E) are not relevant to the T 65 digital ULTRA-TURRAX

®

unit.

Any faults are shown exclusively on the display (F).

Note: If any error codes other than those listed above appear on the display please switch the unit off and on again at the main
switch (A). If the error code persists please contact the IKA

®

Service Department.

If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA

®

service department

- Send the device for repair, including a short description of the fault.

Advertising