Error codes – IKA Dry Block Heater 4 User Manual

Page 15

Advertising
background image

15

Error codes

Any malfunctions during operation will be identified by an error message on the display (LED (Q)).

Proceed as follows in such cases:

Switch off device using the main switch (E)

Carry out corrective measures

Restart device

Error code

Cause

Effect

Solution

E 3

Temperature inside device too high

Heating off

Switch off device and allow to cool down

E 6

Break in safety circuit

Heating off

- Plug in contact plug (D)

- Plug in PT 1000 contact thermometer /

temperature sensor

- Replace faulty connecting cable, plug, or

contact thermometer

E 25

Heating and switching element monitoring

Heating off

- Switch off device

- Warning! Only to be carried out by autho-

rised service personnel: Carry out an internal

test on the device to check the plug-in

connector for the heating element

E 26

Difference between temperature of safety

Heating off

- Switch off device

sensor and temperature of control sensor

- Warning! Only to be carried out by autho-

control temperature > (safety temperature

rised service personnel: Carry out an internal

+ 40 K))

test on the device to check the temperature

sensor


If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA

®

service department,

- send the device for repair, including a short description of the fault.