IKA KS 501 digital User Manual

Page 7

Advertising
background image

7

HS_KS 501 digital_092012

The attachments are placed on the
shaking table and fastened with the
lateral quick-action vises.

Separating funnel attachments are only
suitable for HS 501 digital.

Maintenance and Cleaning

The instrument requires no maintenance. Only the ball bearings are subjected to a
natural wear. If the instrument needs to be repaired, it may only be opened by a
qualified technician. If the instrument does not seem to function, please check the
fuses first: Disconnect mains plug of instrument and open fuse holder. For safety data
see Technical Data/Mains Connection.
The surfaces of the instrument may be cleaned with:
a) water with a tenside containing detergent
b) isopropyl alcohol (e.g. if very soiled) with the exception of the polystyrole attachments.
Please submit for repair only instruments that have been cleaned and are free from
substances dangerous to health.

When ordering spare parts, please indicate type of instrument, serial number

and ordering number of the spare part.

Technical Data

Mains voltage

230 V or 115 V

mains

frequency

50/60Hz

mains connection

GEE-instrument plug with fuses:

230 V = 2 x 0,5 A; 115 V = 2 x 1 A

protection class (DIN 40 050)

IP 21

permissible duration of operation

100%

timer

0-56

min/co

power

input

70

watts

power

output

19

watts

speed

0-300

1/min

speed

display

digital

shaking motion

HS 501

to and fro

KS

501

orbital

max. load incl. attachment

15 kg

permissible ambient temperature

+5°C - +50°C

relative humidity

max. 80%

dimensions W x D x H

505 x 585 x 120 mm

weight (without attachment)

24 kg

Subject to technical changes!

Advertising