Technical data – IKA HB 4 basic User Manual

Page 14

Advertising
background image

14

HB4 0596

HBR4 digital

HB4 digital

HB4 basic

Nominal voltage

VAC

230 ±

10%

230 ±

10%

230 ±

10%

or

VAC

115 ±

10%

115 ±

10%

115 ±

10%

Frequency

Hz

50/60

50/60

50/60

Heating function
Heating output

W

1000

1000

1000

Temperature range (medium)

°C

Room temperature ..200

Room temperature ..200

Room temperature ..225

Temperature range (medium)

Set and actual temperat. LCD Set and actual temperat. LCD Set temperat. graduation

Setting accuracy

K

1

1

±5

Controller deviation

(3 l water / 90 °C)

K

±0,4

±2

±5

Absolute deviation / mean

(3 l water / 90 °C)

K

±1

±2

Controller deviation

(3 l Marlotherm oil / 150 °C)

K

±0,8

±1

Absolute deviation / mean

(3 l Marlotherm oil / 150 °C)

K

±2

±10

Stirring function
Motor

ballbearing brush-free

-------

-------

EC motor

Motor output

W

5

-------

-------

Speed range

rpm

100 - 800, adjustable in

-------

-------

steps of 50 rpm

Minimum filling height

cm

3

3

3

Overvoltage class

II

II

II

Duty cycle

%

100

100

100

Enclosure rating (DIN 40 050)

IP 21

IP 21

IP 42

Ambient temperature

°C

+5 .. 40

+5 .. 40

+5 .. 40

Ambient humidity (rel.)

%

80

80

80

Safety class

DIN 12 877 class 2

DIN 12 877 class 2

DIN 12 877 class 2

Total volume - bath container

litre

5

5

5

Material - bath container

1.4301

1.4301

1.4301

Dimensions (diameter x height)

mm

340 x 250

340 x 250

340 x 250

Weight

kg

4.4

4.0

3.9

Technical data

BA_HB4 30.06.2003 10:07 Uhr Seite 14