Centrifuge 5427 R

English (EN)

31

5.5.3

Temperature monitoring

After the set temperature has been reached, the centrifuge reacts to temperature deviations during
centrifugation as follows:

5.5.4

Start FastTemp

With the FastTemp function, you can immediately start a temperature run without samples, at rotor-specific
or temperature-specific speeds. This will quickly bring the rotor chamber, including rotor and adapter, up
to the set target temperature.

Prerequisites

• The centrifuge is switched on.
• Rotor and rotor lid are correctly mounted.
• The centrifuge lid is closed.
• The temperature and g-force (rcf) or speed (rpm) have been set for the upcoming centrifugation (see

Centrifuging on p. 33).

1. Press the fast temp key.

The display shows

•

FT

•

Actual temperature in the rotor chamber

•

g-force (rcf) or speed (rpm)

The temperature control run FastTemp automatically ends when the target temperature has been
reached. A periodic signal tone sounds.

2. Press the start/stop key to end the temperature control run early.

Deviation from target temperature

Action

±3 °C

Temperature display flashes.

±5 °C

Display shows Err 18. Centrifugation is stopped automatically.

• The centrifuge only stops the run once the rotor has reached the set temperature.

Therefore, there may be a delay of approx. 15 min (rotor S-24-11-AT: 30 min) between the
display indicating the set temperature has been reached and the automatic end of the
temperature control run.