Maintenance
Easypet
®
3
English (EN)
6.2.2
Cleaning the pipette clamp
The pipette clamp assemblies can be replaced, cleaned or autoclaved as described below
(121 °C, 1 bar overpressure for 20 min.)
Aspirating cone
• Can be wiped using a damp cloth
• Can be disinfected with alcohol (ethanol, propanol) or
alcohol-containing disinfectants.
• Repeatedly autoclavable
• Can be replaced
Pipette adapter
• Can be rinsed with demineralized water
• Repeatedly autoclavable
• Can be replaced
Membrane filter
• To be dispose of if contaminated
• Cannot be cleaned
• Can be autoclaved once
• Can be replaced
Sealing
• Can be rinsed with demineralized water
• Repeatedly autoclavable
• Can be replaced
You can also use a membrane filter with a pore size of 0.2 μm.