7 print & export, Print & export – Eppendorf BioSpectrometer fluorescence User Manual
Page 55

55
Methods
Eppendorf BioSpectrometer
®
fluorescence
English (EN)
6.5.7
Print & export
In the last optional method step, you can assemble data packets for all samples of a series of
measurements, or selected samples of a series of measurements, for printing to the printer, export to a USB
stick or export to a PC using a USB cable.
λ grid: 20 nm, min. Δ Abs: 0.200:
The peak is not detected because the predetermined
value for
min. Δ abs is too high. The difference of the
absorbance of the peak and the lowest absorbance in
the grid is less than 0.2 A.
λ grid: 20 nm, min. Δ Abs: 0.050:
The peak is detected.
Select data packets
• Use the cursor keys for navigating and confirm
with
enter.
Softkeys
• [Print]: Start printing.
• [Export]: Start export.
• [Sample]: Select individual sample results.