CellTram® Air · Oil · vario — Operating manual

23

Required equipment

Alcohol (ethanol, isopropanol) or disinfectants containing alcohol

Proceed as follows:

1. Choose the disinfection method which corresponds to the legal regulations and guidelines in

place for your range of application.

2. Wipe all surfaces of the CellTram and accessories using the disinfectant.
3. Flush the pressure tube and universal capillary holder with the disinfectant if required.
4. To remove residual liquid, blow out the pressure tube and universal capillary holder with

compressed air after cleaning.

Hint!

There must not be any residual liquid remaining in the pressure tube or in the universal

capillary holder after cleaning as otherwise steady pressure behavior will be

compromised.