10 troubleshooting, 1 general errors, 11 transport, storage and disposal – Eppendorf epT.I.P.S.Version B User Manual
Page 7: 1 storage, General errors, Storage

Troubleshooting
epT.I.P.S.
®
English (EN)
7
10
Troubleshooting
10.1
General errors
11
Transport, storage and disposal
11.1
Storage
Symptom/
message
Cause
Remedy
Liquid is
dripping from
the tip and/or
the dispensed
volume is
incorrect.
The tip is loose or the pipette tip is
poorly fitted.
Firmly attach the pipette tips.
Use epT.I.P.S. pipette tips.
Pipetting was too quick.
The tip is withdrawn from the
liquid too quickly.
Withdraw the tip slowly from
the liquid with a time delay
(approx. 3 seconds).
Liquid could not
be aspirated.
Filter is blocked.
Carefully remove pipette tip.
Attach the pipette tip again and
retry dispensing.
Eject pipette tip. Use a new
one.
Air temperature
Relative humidity
Atmospheric
pressure
In transport
packaging
-25 °C – 55 °C
10 % – 95 %
70 kPa – 106 kPa
Without transport
packaging
-5 °C – 45 °C
10 % – 95 %
70 kPa – 106 kPa