5 high temperature disinfection, High temperature disinfection – Eppendorf Galaxy CO2 Sensor User Manual
Page 55

Maintenance
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
55
10. Refill the humidity tray (see
Using the humidity tray on p. 30). When you reinstall it, ensure that the
humidity tray is pushed fully back.
11. Leave the incubator on for at least two hours (preferably overnight) to allow conditions to stabilize.
12. When the incubator has stabilized, carry out an Auto-Zero and reprogram the desired CO
2
level. It may
be necessary to open the glass door briefly if, after performing an Auto-Zero, the CO
2
level is too high.
8.5
High temperature disinfection
If your incubator is supplied with the High Temperature Disinfection option, follow the guidelines outlined
for operation, (see
Using the high temperature disinfection (Galaxy R) on p. 68) or (see Using the high
temperature disinfection (Galaxy S) on p. 70) for information on how to correctly and safely operate this
option.