Eppendorf S Series 48S CO2 Incubator User Manual
Page 15

15
New Brunswick Scientific
User’s Guide
Figure 2: Installing Shelves
1 Anti-tip
tab
5. Level the incubator by adjusting the feet. Place a small level on the second shelf of
the incubator. Adjust the leveling feet until the incubator is level and stable. Lock
the leveling legs in place by tightening the locking nuts on each leg.
Install the humidity tray:
1. Install the humidity tray in its location beneath the lowest shelf rack position: insert
the tray diagonally (see Figure 3a), then turn the tray 45
° as shown, until you can
set it in position, resting on the chamber floor, with its front and rear edges resting
just above the shelf rack supports (see Figure 3b).
Figure 3: Installing/Removing the Humidity Tray
a
b
1 Center the tray, making sure it sits within the rack supports.
1
1