4 research pro, Research pro – Eppendorf SOP User Manual

Page 45

Advertising
background image

45

Technical specifications

Eppendorf SOP

English (EN)

9.2.4

Research pro

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits

Error

Systematic

error

Random

error

± %

± µL

± %

± µL

0.5 - 10 μL

light gray
0.5 – 20 μL L
46 mm

1

μL

±2.5 ±0.025

±1.8 ±0.018

5

μL

±1.5 ±0.075

±0.8 ±0.04

10

μL

±1.0 ±0.1 ±0.4 ±0.04

5 - 100 μL

Yellow
2 - 200 μL
53 mm

10

μL

±2.0 ±0.2 ±1.0 ±0.1

50

μL

±1.0 ±0.5 ±0.3 ±0.15

100

μL

±0.8 ±0.8 ±0.2 ±0.2

20 – 300 μL

Orange
20 – 300 μL
55 mm

30

μL

±2.5 ±0.75

±0.7 ±0.21

150

μL

±1.0 ±1.5 ±0.3 ±0.45

300

μL

±0.6 ±1.8 ±0.2 ±0.6

50 - 1 000 μL

Blue
50 - 1 000 μL
71 mm

100

μL

±3.0 ±3.0 ±0.6 ±0.6

500

μL

±1.0 ±5.0 ±0.2 ±1.0

1,000

μL ±0.6 ±6.0 ±0.2 ±2.0

500 – 5 000 μL

Purple
0.1 – 5 mL
120 mm

0.5

mL

±3.0 ±15 ±0.6 ±3.0

2.5 mL

±1.2

±30

±0.25 ±6.25

5.0 mL

±0.6

±30

±0.15 ±7.5

Advertising