Eppendorf Universal Adapter UVette User Manual

Universal adapter uvette® — instructions for use, Universal adapter uvette, Select a wavelength of 550 nm

Advertising
background image

Universal Adapter UVette® — Instructions for Use

Instructions for Use

\ \m\

Universal-Adapter UVette®

See

on page

Fig.

Tab.

S.

Leaflet

1

Universal Adapter UVette

®

To enable the Universal Adapter to be used with UVette

®

in as many spectro-

photometers as possible, it is first necessary to determine the light center height of the

spectrophotometer.
If no specifications for the light center height are available from the

manufacturer, it is possible to determine this height using the

measuring gauge:

Select a wavelength of 550 nm.

Insert the measuring gauge vertically into the cuvette holder (in its

original condition) until it reaches the bottom.

Read the height of the light spot.

The most common light center height is 15 mm, which enables the

adapter to be used without being altered.

Advertising