Galaxy® co2 sensor setup — instructions – Eppendorf Galaxy Gas Analyzer User Manual
Page 30

Galaxy® CO2 Sensor Setup — Instructions
30
2. Rear cover removal:
48 S – Remove the CO
2
Sensor access panel by unscrewing the 4 x Pozidriv head screws
with washers (see all screws in dashed red circles in top left-hand image below).
14 S – Remove the rear access panel by unscrewing the 6 x Pozidriv head screws with
washers (see all screws in dashed red circles in top right-hand image below). The top two
screws secure the top cover.
3. With the access panel removed, locate the insulation block covering the CO
2
Sensor and
carefully remove the block (shown in yellow dashed boxes in images below) and place to one
side for refitting later. The rear of the CO
2
Sensor will now be visible.
4. Remove the existing CO
2
Sensor by disconnecting the 5 way Molex connector and then
unscrew the 3 x M3 securing nuts and place all the fixings to the side for refitting later (only
one nylon washer will be reused).
3
Galaxy 48 S / 14 S 120 V & 230 V CO
2
In
cubators