Test procedures – Bose 802c User Manual
Page 10

10
TEST PROCEDURES
4.4 Apply a 100Hz, 500mVrms signal to
the input of the 802C.
4.5 Reference a dB meter to the 302 AMP
output jacks (while in the Bi-Amp mode).
4.6 Refer to the frequency response table
below and verify the response of the unit.
6. Noise Test
6.1 All noise measurements are ANSI
A-weighted true rms, with the inputs
shorted.
6.2 Refer to the table below for the proper
noise levels.
5. Distortion Test
Note: Total harmonic distortion must be
measured in all modes to assure proper
operation.
5.1 Apply a 750Hz, 5Vrms signal to the
input of the 802C.
5.2 Refer to the table below for the distor-
tion specification for the corresponding
mode being tested.
Note: For the Bi-Amp 302 reading the
input signal should be a 100Hz, 5Vrms
signal.
Frequency
Output
Tolerance
55Hz
+3.7dB
±
1.5dB
100Hz
0dB
(reference)
-
230Hz
-8.3dB
±
1.5dB
750Hz
-25.8dB
±
2.5dB
302 Bi-Amp Frequency Response Table
Output
Frequency. Distortion
802 Full
Range
750Hz
<0.1%
302 Passive
750Hz
<0.1%
802 Bi-Amp
750Hz
<0.1%
302 Bi-Amp
100Hz
<0.1%
Distortion Table
Noise Table
Output
Noise
802 Full
Range
<20uV
302 Passive
<20uV
802 Bi-Amp
<20uV
302 Bi-Amp
<10uV