5

www.cbsscientific.com

SECTION 1

General Information
1.1 Introduction

Horizontal Mini-Gel systems (GC)MGU-102T and (GC)MGU-202T provide a compact
format to rapidly separate and identify DNA and RNA fragments, PCR products and
synthetic oligonucleotides, while using minimal quantities of expensive reagents. The
tapered baffle system allows gel casting directly onto the unit or removable trays
without having to seal with agarose plugs or tape. The gel bed and gel tray are UV
transparent. Combs are offered in several thicknesses and formats to accommodate
a wide variety of experimental applications.

1.2

Specifications

Constructions

Buffer chamber, lid

Acrylic

Electrodes

Platinum wire .010” diameter

Power cords

FR Silicone rated 7500VDC, 200mA, 65ºC

Tapered baffles

Polycarbonate

Combs

Polycarbonate, Anodized Aluminum or Teflon

Comb backing

Acrylic

UVT Gel Tray (optional)

U-V transparent acrylic

Safety Certification

EN61010-1-1993 (IEC1010-1)

*Calculated using gel thickness of 0.5cm

1.3

Safety

Power to the Mini-Horizontal units is to be supplied by an external DC voltage power
supply that must be ground isolated so that the DC voltage output floats with respect
to ground. For any power supply used, the maximum specified operating parameters
for the units are:

Maximum Limits

250VDC voltage

15 watts power

60 mA current

50*C ambient temperature

Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed. Do not attempt to use the unit without the safety
lid. Always turn the power supply off before removing the lid, or when working with
the unit in any way. Follow safety precautions specified by the power supply manu-
facturer.

Model #

(GC)MGU-102T

(GC)MGU-202T

Shipping weight

3 lb.

3 lb.

Overall Size W x L x H (cm)

14 x 19.7 x 12.7

17 x 19.7 x 12.7

Gel Bed Size W x L (cm)

5.5 x 8.5

7.5 x 10

Recommended Buffer Volume

150ml

250ml

Agarose Solution Required*

23ml

36ml

# of Slots in Gel Tray

2

2

Pathlength/Slot (cm)

5.0

5.5

Voltage limit

250VDC

250VDC