10
E7
Short Circuit
Load current exceeded 2 A
1. Check for any short circuit or evidence
of electrical arching.
2. Buffer concentration is too high.
3. Press any button* to clear the error
code.
*Press any button, except the RUN button, to clear the error message.
D-3. Replacing the fuses
1. Unplug the power cord from the power supply.
2. Locate the fuse compartment on the rear panel.
3. Use a screwdriver to pry out the fuse compartment. There are two fuses in the compartment.
Check and replace any blown fuses with a fuse of the same size and ratings.
4. FUSES (2); 250 V, 3.15 A, 5 x 20 mm fast blow.