Carbolite 2116 User Manual
Page 7

2132/2116 Controller
MC02 – 1.08 7
3.4
Stopping the Timer
To stop the timer at any time while it is running, change the
StAt
User Calibration parameter to
OFF
. This is the same as reducing tmr to zero. The controller then acts as though at has reached
the end of the time period.
3.5
End of Time Period
Modes 1 and 3: heating stops at the end of timing; the
m-A
parameter changes to
mAn
.
Modes 2 and 4: heating continues at the end of timing; the
m-A
parameter remains at
Auto
.
Mode 5: heating starts at the end of the timing period; the
m-A
parameter remains at
Auto
.
In modes 1 to 4 the alarm message
EnD
flashes on the display at the end of timing; the
StAt
parameter remains at
run
.
In mode 5 there is no End message; the
StAt
parameter changes to
OFF
at the end of timing.
3.6
Cancelling the Alarm
To acknowledge (cancel) the
EnD
alarm, press Page and Scroll together; the
StAt
parameter
changes to
OFF
.
Alternatively cancel the alarm by directly changing the
StAt
parameter from
run
to
OFF
.
3.7
Program Example
To heat up at 10°C per minute to 500°C; to hold at 500°C for 1 hour; then to allow to cool down.
(This example uses timing mode 1, as on the first row of the table on page 5, but also includes
the use of ramp rate.)
to create this program
1. Start with display at “home”; use arrow keys to alter the setpoint to
500
.
2. Press scroll until
sp.rr
shows; use arrow key to set value to
10
(if you do not want to limit the ramp rate, ignore this step or set the value to
OFF
)
3. Press scroll until
tm.op
shows; use arrow key to set value to
opt.1
4. Press scroll until
dwel
shows; use arrow key to set value to
60
to run this program
5. Press scroll until
stat
shows; use arrow key to set value to
run
- heating starts when
run
is set;
- timing starts when the working setpoint reaches 499°C;
- heating stops 61* minutes later, and
End
flashes on the display.
6. Cancel
End
by pressing page and scroll together.
* note that the timer runs for an extra minute, when
tmr
has counted down to zero; allow for this
when testing the system with short durations such as 1 or 2 minutes..