APPENDIX B –LIGHT INTENSITY CHARACTERISTICS (REFERENCE ONLY)

Light intensity nomenclature

Near UV

Visual

Abbreviation

UVA

VIS

Lamp type

Black

Cool white

Wavelength

320 – 400 nm

400 – 800 nm

Intensity units

W/m

2

lux (or klux)

Exposure (dose) units

W-hr/m

2

lux-hr (or klux-hr)

ICH Q1B confirmatory requirement

200 W-hr/m

2

1.2 million lux-hr

or 1200 klux-hr

Photopic (visual light) conversions

1 klux =

92.9 foot-candles (fc)

13.5 μmol/m

2

/s (for cool white fluorescent only)

Light intensity and lamp age

The life expectancy of a fluorescent lamp varies significantly by the operating condition in which it is used. Under

ideal conditions, maximum light intensity decreases with lamp age as shown below. Other factors influencing lamp

life includes chamber air temperatures (optimal at 25ºC), number of on/off cycles, and amount of light dimming

applied. For optimal performance, CARON recommends lamp replacement every 2000 hours.

Optimal Fluorescent Lamp Output

80

85

90

95

100

0

1000

2000

3000

4000

5000

Operating Hours