Caron 6404-4 User Manual
Page 33
Advertising

6400/6404 Series Operations Manual
Page 33 of 54
Rev F 09/18/14
Setting High/Low O2 Alarm Level (suppressed O2 model 6404s only)
The incubator has both a high and low programmable O2 alarm. The factory default for
the high O2 alarm is 1.0% above O2 setpoint. The factory default for the low
Temperature alarm is 1.0% below O2 setpoint. To change either alarm level:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select High O2 Alarm / Low O2 Alarm
5) Press the key to enter High O2 Alarm / Low O2 Alarm
6) Press the to change alarm level
7) Press the key to save the alarm level
8) Press the key twice to return to the main menu
Advertising