Caron 6017 User Manual

Page 36

Advertising
background image

6010 Series Operations Manual

Rev E 2/14/2012

Page 36 of 68

Controlling the Diurnal Lights (6012) and Optional Fluorescent Lights (LGHT303)

The diurnal chamber comes with a fluorescent lighting system used to simulate day and

night testing. This option is also available on the model 6010 & 6015 as LGHT303.

There are separate temperature and humidity (optional) set-points that correspond with

the lights on (day) and lights off (night). The set-points can also be made the same if

continuous conditions are needed throughout the light cycles.

The diurnal lighting system consists of two separate light banks. Each light bank has

two lamps and is mounted on the underside of a shelf. The banks can be located on the

same or separate shelves. Each light bank has a corresponding switch that enables

that light bank to come on when the timer is on. Only the light banks that have their

individual switch turned on will illuminate. The green indicator light above the light bank

switch signifies when each light bank is on.

For units equipped with ramping controllers (DLUX301 & DLUX302), the timer can be

used to trigger an event in a profile to switch between two set-points (if programmed

accordingly).

Individual

Light Bank

switches

Light Power

Switch

Light Bank

Indicators

Light Cycle

Timer

Advertising
This manual is related to the following products: