Page 11-2

20638-02 Rev 01

•

‘

→Sensor Serial #’ will appear on Line 2 of the display. Press the

‘ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys (

↑ and ↓ to scroll through the values and

← and → to change position) to enter the sensor serial number.
Once all digits are entered, press the ‘ENTER’ key to save to
memory. ‘

→Sensor Serial #’ will again be displayed on Line 2.

•

Next, press the

↓ key to scroll to the ‘Pipe Size’ menu on Line 2.

Press the ‘ENTER’ key to access the options available under that
menu. Note: it is necessary to access only one of the following
options.

•

The first choice on the ‘Pipe Size’ menu is ‘

→ID / Wall’ shown on

Line 3 of the display. If this value is known it can be entered here.
Press the ‘ENTER’ key and the current saved inner diameter and
wall thickness will be displayed on Line 4 of the display. Use the
arrow keys (

↑ and ↓ to scroll through the values and ← and → to

change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
‘ENTER’ key to save the value to memory.

•

The second choice on the ‘Pipe Size’ menu is ‘Size/Sched’ (pipe
Size / Schedule). If this value is known press the ‘ENTER’ key.
Use the arrow keys to enter the values and then press the
‘ENTER’ key to save to memory.

•

The third choice on the ‘Pipe Size’ menu is ‘OD / Wall’ displayed
on Line 3. Use the arrow keys (

↑ and ↓ to scroll through the values

and

← and → to change character) to enter the pipe inner

diameter. Units can be inches or millimeters. Once all digits are
entered, press the ‘ENTER’ key to save the value to memory.

•

Following the entering of ‘Pipe Size’ press the ‘BACK’ key and
‘

→Pipe Size’ will be displayed on Line 2. At this point the user can

either press the ‘ENTER’ key to re-enter the ‘Pipe Size’ menu or
press the

↓ key to move to the next menu item.

•

Pressing the

↓ key will next display ‘→Fluid Properties’ on Line 2.

Press the ‘ENTER’ key and ‘

→Specific Gravity’ will appear on

Line 3 and the current value on Line 4. To change the Line 4 value
press ‘ENTER’ and use the arrow keys to enter the new value.
Water at 25 ºC (0.997) is the default. Appendix E lists values for
water at various temperatures. Once the new value has been
entered on Line 4, press ‘ENTER’.

•

If no change is made (or after a change in Specific Gravity has
been made) pressing the

↓ key will next display ‘→VISCOSITY (Pa

s)

’ on Line 3 and the current value on Line 4. To change the Line

4 value press ‘ENTER’ and use the arrow keys to enter the new
value. Water at 25 ºC (8.9008 e

-04

) is the default. Appendix E lists