Calibration Procedure - KM820/V

Verifying Infra-Red Non contact thermometers

•

Insert the reference thermometer probe (e.g.

KM20Ref) fully into one of the holes in the Validator.

•

Allow the instruments to stabilise to environment

conditions.

•

Switch on both instruments after stabilisation

period.

•

Point the Infra-Red thermometer at the base of the

Validator while ensuring the nozzle is as far inside

the Validator cup as is possible. Note the variation.

(acceptable error 2.5°C)

Verifying Contact Thermometers

•

Insert the reference thermometer probe (e.g.

KM20Ref) fully into one of the holes in the Validator.

•

Insert the probe of the Instrument being verified

fully into the other hole in the Validator.

•

Allow the instruments to stabilise to environment

conditions.

•

Switch on both instruments after stabilisation

period.

•

Note the variation. The acceptance error will be

dependent on the specification of the unit being

verified.

Note: The time for the instruments to stabilise will be dependent on

the validator reaching its environment temperature, which at freezer

temperatures could be up to 2 hours.

Care must be exercised when handling the validator after

exposure to low temperatures.