351 / 352 / 353 SPECIFICATIONS

351 / 352 / 353 TROUBLESHOOTING

TEMPERATURE RANGES:
35100-K / 35200-K / 35300-K:

-100° to 999°F (-73° to 537°C)

35132 / 35135 / 35140 / 35141:
35232 / 35235 / 35240 / 35241:
35332 / 35335 / 35340 / 35341:

-100° to 500°F (-73° to 260°C)

ACCURACY:
35100-K / 35200-K / 35300-K:

With factory calibration, ±0.5F° (±0.3C°) over entire
measurement range at ambient temperatures
between 68° to 86°F (20° to 30°C)

35132 / 35140 / 35141:
35232 / 35240 / 35241:
35332 / 35340 / 35341:

With factory calibration, ±0.9F° (±0.5C°) Total
System Accuracy over entire measurement range
at ambient temperatures between 68° to 86°F
(20° to 30°C)

35135 / 35235 / 35335:

Surface probe temperature errors for draft-free, flat,
clean oiled surfaces with 2.2 lbs.(1 Kg) of pressure
are typically within +2.7°F and –6.3°F (+1.5°C and
–3.5°C) without thermostat recycling.

RESOLUTION:
35100-K / 35300-K:
35132 / 35135 / 35140 / 35141:
35332 / 35335 / 35340 / 35341:

Tenth degree (.1)

35200-K / 35232 / 35235 / 35240 / 35241:

Tenth degree / whole degree (.1/1), user selectable

BATTERY LIFE:

1800 Hours (2) AAA Alkaline batteries

AMBIENT TEMPERATURE LIMITS:

32° TO 122°F (0° TO 50°C)

TRACEABILITY:

We certify that this instrument has been calibrated
to temperature and/or electrical standards traceable
to the U.S. Department of Commerce National
Institute of Standards and Technology.

CALIBRATION:

Under normal operating conditions, the unit should
not require re-calibration.

TROUBLESHOOTING

Our Customer Service Department is available
Monday through Friday 8:00 AM to 5:00 PM EST
at 800-835-5011.

1) Unit Does Not Turn On:

Batteries are installed incorrectly or dead.
Check batteries are installed according to picture on
rear label. Replace batteries.

2) Display Shows “Prob”: Probe is broken (Open

thermocouple see “Test for Broken Probe” below).

3) Test for Broken Probe: Pull the probe connector out

partially to expose the blade contacts. Place a metal
paper clip across connector blades to short contacts. If
instrument displays temperature, the probe is broken.
Replace Probe.

4) Battery Icon Appears on Display: Low battery.

Replace batteries.

5) Display Shows “bAtt” and Instrument Shuts Off:

Replace batteries.

6) Display Shows “LO” or “HI”: Temperature measured

is outside measurement range of instrument.

351 / 352 / 353 WARRANTY

WARRANTY: Cooper-Atkins thermometers are backed
by a 5-year Instrument Warranty and 1-year Probe
Warranty on defects in workmanship and materials.