background image

13

NAME/CODE

DESCRIPTION

CODE/VALUE

A1
Baud Rate

Selects the baud rate for data transmission through the

serial port.

1200
2400 4800
9600√
19200

A2
Data Bits and
Parity

Selects the number of data bits and parity of serial

transmission. "8n"=8 data bits with no

parity bit and one stop bit.

"7O"=7 data bits with odd parity bit and one stop bit.

"7E"=7 data bits with even parity bit and one stop bit.

"7n"=7 data bits with no parity bit and two stop bits

8n√

7O

7E

7n

A3
Mode of
Serial
Transmission

Selects when data will be sent out of the serial port to a

printer or computer.

"C" = Continuous mode; send data continuously

"d" = Demand mode; send data when a PRINT

command is issued from the printer, computer, or

indicator.

C

d√

A4
Display
Check

Actuates the function that illuminates all digit segments,

decimal points, and LCD annunciators in a test

sequence. Pressing the ZERO key to scroll down one

level begins the test sequence.

Press ZERO key

to begin

sequence

A5
Disable the
lb/kg Key

Allows the lb/kg key to be disabled so that an operator

cannot accidentally press the key and change the

display units.

"0" = Disable the lb/kg key "1" = Enable the lb/kg

key

0
1√