Heidolph ROTAVAC 20 User Manual

Page 30

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 01-005-005-36-1

page 30 of 35

Translation of the original instructions

6.5 Replacing the overpressure safety relief device at the condenser

Overpressure safety relief device
...................................................................... 23-30-01-04-98

Round bottom flask 500 ml, coated
...................................................................... ...514-83000-02

➨ Remove joint clip at the catchpot.

➨

Unscrew the four Torx screws at the counter holder of the

condenser and remove condenser. Thereby remove the
adapter from the inlet of the condenser.

➨ Pull off old overpressure safety relief device and install

new one. Check for correct position of the PTFE-foil under
the overpressure safety relief device.

➨ Reassembly in reverse order.

➨ Position adapter in the inlet of the condenser and screw

condenser with counter holder to the pump (Torx screws).
Fix with union nut.

➨ Reassemble catchpot.

Advertising