Disposal, Troubleshooting, Specifications – Heidolph REAX 2 User Manual

Page 18

Advertising
background image

18

E

Caution
Do not jolt or shake the item during transport.

DISPOSAL

For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).

TROUBLESHOOTING


Work on electric and electronic components is limited to qualified personnel.

Master switch on shaker / mixer won't light
1. check power cord
2. check circuit breakers

Item won't shake (master switch lighted))
1. Thermal motor circuit breaker triggered by motor overload

Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.

2. Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).

SPECIFICATIONS


REAX 2

power supply

230/240 V 50/60 Hz / 115 V 50/60 Hz (optional)

input power

27 W

shaker frequency

20

– 100 1/min

mixing movement

overhead rotation

drive

split-pole motor, features electronic speed control;
motor overheating protection with automatic reset

adapter

Interchangeable, see accessories

Load-bearing capacity

max. 1 kg, limitations for off-center load

weight

4,8 kg

protection class

IP 21

ambient temperature

0°C to 50°C at 80% rel. humidity
Approved for installation in gassing and conditioning
cabinets (make reference to temperature limits)

space required

460 mm x 175 mm