Designation

PharMed

Silicone

Viton

Advantages

Ideal for non-toxic and non-

haemolytic cell cultures and

tissues. Impervious to normal

light and UV radiation. The tube

can be welded, bonded and

formed. Very long life. Low gas

permeability.

Excellent bio-compatibility.

Practically no disassociation of

plasticizers and additives.

Neutral to smell and taste. Not

susceptible to mildew. Ideal for

low temperatures. Waterproof;

resistant to ozone, radiation and

sunlight. Does not warp. Silicon

peroxide; hot vulcanized.

Highly resistant to chemicals.

Excellent resistance to corrosive

media, solvents and oils at high

temperatures. Only slightly

pervious to gas.

Limitations

Additives may become

disassociated.

Unsuitable for concentrated

acids, solvents, oils, acids or

diluted caustic soda. Relatively

highly pervious to gas.

Limited life span.

Use with:

Acids

Lyes

Solvents

Pressure

Vacuum

Viscous media

Sterile media

good

good

unsuitable

good

excellent

good

excellent

under certain conditions

under certain conditions

unsuitable

satisfactory

good

satisfactory

excellent

excellent

excellent

varies, tests recommended

good

good

good

satisfactory

Temperature range

-60 to +130 °C

-50 to +230 °C

-30 to +200 °C

Physical properties

Thermoplastic elastomeron a

polypropylene basis with

plasticizers. Excellent tensile

strength. Opaque, beige.

Polydimethylsiloxane with sili-

ceous earth and silicone

additives. Translucent, white.

Excellent resistance to contact

pressure.

Fluorocarbon rubber, thermally

formed.

Viton B (67 fluorinated).

Opaque, black.

Complies with

standards

USP, Class VI

FDA21 CFR 177.2600

NSF (standard 51)

USP, Class VI

FDA21 CFR 177.2600

Permeability CO

2

H

2

O

2

N

2

450

-

75

29

20132

6579

7961

2763

76-79

-

13-15

4,3

Cleaning/sterilization

Can be autoclaved without

ageing symptoms.

Can be cleaned with hot water

and suds. Use only soap

without oil additives; use no

synthetic detergents. Rinse

with distilled water. Can be

sterilized by autoclaving in a

moist atmosphere or with

gamma rays. Cannot be

sterilized with ethylene oxide.

Can be sterilized at 249°C in

16 hours in a circulating air

heating cabinet.

®

®